Location -> Products -> Analyzer Instrument -> On-line Instrument -> Industry Dissolved Oxygen Analyzer

Type:CX-IDO-1304A industrial dissolved oxygen meter
Illustrate:It is a special instrument for measuring oxygen content in the liquid in boiler feed water, condensate water, environmental protection sewage and other industries.
CX-IDO-1304A Industrial Dissolved Oxygen Meter
Brief Introduction
1.CX-IDO-1304A Industrial Dissolved Oxygen Meter is a high-intelligence on-line continuous monitor made by our company. 2.It can be equipped with CX-IDO-130F polar organphic electrode, achieving the wide-range measurement from ppb level to ppm level automatically
3.Relying on its quick response, stability, reliability, low usage cost and other features, CX-IDO-130F Dissolved Oxygen Electrode is suitable for the large-scale use by the customers.
Main Features
1.Computer-based: It adopts high-performance CPU chip, high-precision AD conversion technology and SMT to complete multi-parameter measurement, temperature compensation and automatic range switching, with high precision and good repeatability.
2.Reliability: Touch button, no switch knob or potentiometer
3.Automatic conversion of measuring frequency: The electrode polarization is avoided and thus the measurement accuracy is improved.
4.Alarm: Alarm signal isolation output; the alarm upper and lower limits can be arbitrarily set and the alarm isundone in hysteresis.
5.Network function: Isolated current output and DRS485 or RS232 communication interface.
6.Isolated current output: The current is corresponding to the output upper and lower limits of the dissolved oxygen values, which can be arbitrarily set.
Main Parameters
Measuring range | 0~100.0 ug/L; 0~20.00 mg/L(automatic switching); 0~60°C |
Resolution | 0.1 ug / L; 0.01 mg / L; 0.1°C |
Intrinsic error of the whole instrument | ug/L: ± l.0%FS; mg/L: ±0.5%FS, temperature: ±0.5°C |
Repeatability of the indication | Stability of the indication of the whole instrument: ±l.0%FS |
Automatic temperature compensation range | 0~60°C, with 25°C as the reference temperature. |
Response time | <60s (98% and 25°C of the final value) |
Output current error | ≤±1.0%FS |
Isolated output | 4~20 mA(load resistance <750Ω) |
Communication interface | RS485 or 232(optional) |
Data storage time | 10 years |
Alarm relay | AC 220V, 3A |
Power supply | 220V±10%50 ±1 Hz |
Overall dimension | 96 (length)x96(width)×110(depth)mm, Dimension of the hole: 92x92 mm |
Weight | 0.8kg(secondary meter) |
Working conditions | ambient temperature: 0-60°C; relative humidity <85% |
Connection tubes for inlet and outlet water | Φ8, Φ10 and Φl2 |
Last:CX-IDO-1304 industrial dissolve.. Next:CX-IDO-1304B industrial dissolv..