Type:CX-IDO-1305 Industrial Dissolved Oxygen Meter
Illustrate:CX-IDO-1305 Industrial Dissolved Oxygen Meter is a kind of high-intelligence in-line meter that is developed by our company and used for continuous on-line monitoring. It can be equipped with DOG-208F Polarographic Electrode and can automatically switch from ppb level to ppm level of wide-range measurement.
CX-IDO-1305 Industrial Dissolved Oxygen Meter
Brief IntroductionCX-IDO-1305 Industrial Dissolved Oxygen Meter is a kind of high-intelligence in-line meter that is developed by our company and used for continuous on-line monitoring. It can be equipped with DOG-208F Polarographic Electrode and can automatically switch from ppb level to ppm level of wide-range measurement.
Main Features
1.All the data are displayed in English. It can be easily operated
2.It has a complete English display and elegant interface.
3.Liquid crystal display module with high resolution is adopted.
4.All the data, status and operation prompts are displayed in English. There is no symbol or code that is defined by the manufacturer.
5.Simple menu structure and text-type man-instrument interaction: Compared with traditional instruments
6.Muiti function: it adopts classified menu structure, which is similar to that of a computer, it is clearer and more convenient. It is not necessary to remember the operation procedures and sequences.
7.It can be operated according to the prompts on the screen without the guidance of an operation manual.
8.Multi-parameter display: The oxygen concentration value, input current (or output current), temperature values,
time and status can be displayed on the screen at the same time.
9.The main display can show the oxygen concentration value in a 10 x 10mm size.
10.As the main display is eye-catching, the displayed values can be seen from a long distance away.
11.Six sub-displays can display such information as the input or output current, temperature, status, week, year, day, hour, minute and second, so as to adapt to different users’ habits and to conform with different reference times set by the users.
Technical Parameters
Measuring range | 0~100.0ug/L; 0~20.00 mg/L (automatic switching); (0-60°C);(0-150°C)Option |
Resolution | 0.1ug/L; 0.01 mg/L; 0.1°C |
Intrinsic error | ug/L: ±l.0%FS; mg/L: ±0.5%FS, temperature: ±0.5°C |
Repeatability | ±0.5%FS |
Stability | ±1.0%FS |
Automatic temperature compensation range | 0~60°C, with 25°C as the reference temperature. |
Response time | <60s (98% and 25°C of the final value) 37°C: 98% of the final value < 20 s |
Clock accuracy | ±1 minute/month |
Output current error | ≤±l.0%FS |
Isolated output | 0-10mA (load resistance <15KΩ), 4-20mA (load resistance <750Ω) |
Communication interface | RS485 (optional) (Double power for option) |
Data storage capacity | l month (1 point/5 minutes) |
Saving time of data under a continuous power-failure condition | 10 years |
Alarm relay | AC 220V, 3A |
Power supply | 220V±10% 50±1HZ, 24VDC(option) |
Protection | IP54, Aluminium she |
Size |
secondary meter: 146 (length) x 146 (width) x 150(depth) mm; dimension of the hole: 138 x 138mm |
Working conditions | ambient temperature :0-60°C; relative humidity <85% |
The connection tubes for inlet and outlet water: Pipes and hose |